An Extremely Inhomogeneous Gross Tumor Dose is Suitable for Volumetric Modulated Arc-Based Radiosurgery with a 5-mm Leaf-Width Multileaf Collimator for Single Brain Metastasis

Search this article

Description

Introduction Single or multi-fraction (mf) stereotactic radiosurgery (SRS) is an indispensable treatment option for brain metastases (BMs). The integration of volumetric modulated arc therapy (VMAT) into linac-based SRS is expected to further enhance efficacy and safety and to expand the indications for the challenging type of BMs. However, the optimal treatment design and relevant optimization method for volumetric modulated arc-based radiosurgery (VMARS) remain unestablished with substantial inter-institutional differences. Therefore, this study was conducted to determine the optimal dose distribution suitable for VMARS of BMs, especially regarding dose inhomogeneity of the gross tumor volume (GTV). The GTV boundary, not margin-added planning target volume, was regarded as a basis for planning optimization and dose prescription. Materials and methods This was a planning study for the clinical scenario of a single BM. Eight sphere-shaped objects with diameters of 5-40 mm in 5-mm increments were assumed as GTVs. The treatment system included a 5-mm leaf width multileaf collimator (MLC) Agility® (Elekta AB, Stockholm, Sweden) and a dedicated planning system Monaco® (Elekta AB). The prescribed dose (PD) was uniformly assigned to just cover 98% of the GTV (D

Journal

  • Cureus

    Cureus 15 2023-02-25

    Springer Science and Business Media LLC

Citations (10)*help

See more

References(30)*help

See more

Related Projects

See more

Keywords

Report a problem

Back to top