石灰化組織のマイクロCT画像におけるアーチファクト低減法

書誌事項

タイトル別名
  • Artifact Reduction Method for Micro-CT Image in Mineralized Hard Tissue

この論文をさがす

抄録

The aim of this study was to reduce artifacts in microcomputed tomography (micro-CT) images caused by the beam hardening effects of X-ray. The micro-CT images were corrected in comparison with a human extracted tooth embedded in compensating material, dental plaster (CaSO4・1/2H2O), which was placed in an aluminum cylindrical container and not embedded. As a result, the artifacts of the micro-CT images were reduced by the compensating materials. It is considered that this phenomenon occurred because the low-energy X-ray parts were absorbed by the compensating material which has a high attenuation coefficient. The development of this method eliminated the problems caused by beam hardening effects when measuring bone mineral density (BMD), such as surrounding newly formed bone in dental implants, and also suggested the possibility of using the new methods for BMD by using micro-CT images.

収録刊行物

参考文献 (9)*注記

もっと見る

関連プロジェクト

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ