Evaluation of a High-resolution Gas Chromatography/High-resolution Mass Spectrometry Method for the Determination of Dioxin and Related Compounds.
-
- TAKASUGA Takumi
- Shimadzu Techno-Research Inc.
-
- INOUE Tsuyoshi
- Shimadzu Techno-Research Inc.
-
- OHI Etsumasa
- Shimadzu Techno-Research Inc.
Bibliographic Information
- Other Title
-
- ダイオキシンおよび関連化合物質のHRGC/HRMSによる超微量分離定量分析における最適条件化の検討
Search this article
Description
Dioxin analysis requires high sensitivity, high selectivity and specificity. Currently, the analytical method of choice is high-resolution gas chromatography/high-resolution mass spectrometry (HRGC/HRMS) using the SIM method after extensive clean-up.<BR>In this paper, relative efficiencies and performances of both of various sample clean-up methods and HRGC/HRMS method were investigated and evaluated in detail. Some of the noteworthy problems were pointed out and discussed, and the most suitable combination of the procedures for Dioxin analysis were suggested.
Journal
-
- Journal of Environmental Chemistry
-
Journal of Environmental Chemistry 2 (3), 599-613, 1992
Japan Society for Environmental Chemistry