Calibration of a Superconducting Quantum Interference Device Magnetometer
-
- Koyano Tamotsu
- Cryogenics Center, University of Tsukuba
Search this article
Abstract
Calibration factor for magnetization scale depends on sample size along the scan direction in a superconducting quantum interference device (SQUID) magnetometer. A reference material with the same shape and size as a sample or the proper correction of apparent values is necessary for the accurate determination of magnetization.
Journal
-
- Japanese Journal of Applied Physics
-
Japanese Journal of Applied Physics 43 (10), 7322-7323, 2004
The Japan Society of Applied Physics
- Tweet
Details 詳細情報について
-
- CRID
- 1390282681240346624
-
- NII Article ID
- 10013746356
- 210000056603
- 130004531438
-
- NII Book ID
- AA10457675
-
- ISSN
- 13474065
- 00214922
- http://id.crossref.org/issn/00214922
-
- Text Lang
- en
-
- Data Source
-
- JaLC
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed