- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Knowledge Graph Search feature is available on CiNii Labs
- 【Updated on June 30, 2025】Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
Studies of Artifacts with TFE Factor Increase in Heart Delayed Enhancement MRI Sequence IR-T1TFE
-
- HORIE TOMOHIKO
- Department of Radiology, Tokai University Hospital
-
- IKEDA SYU
- Department of Radiology, Tokai University Hospital
-
- MURO ISAO
- Department of Radiology, Tokai University Hospital
Bibliographic Information
- Other Title
-
- 心臓遅延造影MRI Sequence IR-T1TFEにおけるTFE Factorの増加に伴うArtifactの検討
Search this article
Description
The characteristic of delayed enhancement MRI is high spatial resolution, which makes it possible to evaluate the degree of damage to the myocardium from the inner to the outer membrane of patients with ischemic heart disease. Therefore, this MRI technique is unique in its ability to detect myocardial condition and is necessary to obtain high-quality images. We experienced artifacts induced by TFE factor increase with delayed enhancement of IR-T1TFE. The purpose of this study was to determine the cause of such artifacts. IR-T1TFE changed signal intensity with phase direction as the TFE factor increased. Streak artifacts occurred because signal intensity caused changes with phase direction. Increases in TFE factor prolonged data collection time, such that marked artifacts were created because of changes in signal intensity. Ghost artifacts occurred because signal intensity changed between shots. When the TFE factor was increased, the difference in signal intensity was diminished between shots. The interval of acquired noise decreased in the raw data. Therefore, the interval of ghost artifacts became wider on images.
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 60 (5), 708-718, 2004
Japanese Society of Radiological Technology
- Tweet
Details 詳細情報について
-
- CRID
- 1390282681365587712
-
- NII Article ID
- 110003434609
-
- NII Book ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
- http://id.crossref.org/issn/03694305
-
- PubMed
- 15213699
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- Crossref
- CiNii Articles
- OpenAIRE
-
- Abstract License Flag
- Disallowed