- 【Updated on January 20, 2026】 Integration of CiNii Books into CiNii Research
- Trial version of CiNii Research Knowledge Graph Search feature is available on CiNii Labs
- 【Updated on November 26, 2025】Regarding the recording of “Research Data” and “Evidence Data”
- CiNii Research researchers search function has been released.
Reference Values for the Automatic Screening of QT Prolongation in School Cardiac Examinations
-
- Hazeki Daisuke
- Kagoshima City Hospital, Department of Pediatrics Kagoshima City Cardiac Screening Committee
-
- Shiokawa Naohiro
- Kagoshima City Hospital, Department of Pediatrics Kagoshima City Cardiac Screening Committee
-
- Ueno Kentarou
- Kagoshima City Cardiac Screening Committee Kagoshima University Hospital, Department of Pediatric
-
- Kusubae Ryo
- Kagoshima City Hospital, Department of Pediatrics Kagoshima City Cardiac Screening Committee
-
- Nomura Yuichi
- Kagoshima City Hospital, Department of Pediatrics Kagoshima City Cardiac Screening Committee
-
- Yoshinaga Masao
- Kagoshima City Cardiac Screening Committee National Hospital Organization Kagoshima Medical Center, Department of Pediatrics
Bibliographic Information
- Other Title
-
- 学校心臓検診におけるQT延長スクリーニングを自動計測値で行うための抽出基準値に関する検討
Search this article
Description
<p>Background: The heart-rate–corrected QT interval (QTc) by automatic measurements is longer than that by manual measurements. The reference QTc values used for screening QT prolongation in school cardiac screening are generated by manual measurements; when such values are used in automatic measurements, overscreening occurs.</p><p>Methods: We retrospectively analyzed the electrocardiogram records of all first- and seventh-grade students who underwent school cardiac examinations in Kagoshima City from 2009 to 2013. Only students screened using the automated QTc values underwent manual measurements, and those who exceeded the manual reference QTc values were selected. The automated reference values were developed according to the automatically measured QTc values of the extracted students.</p><p>Results: Out of 54,586 students, 1233 (2.3%) were screened using manual measurement. Among them, 52 (0.10%) exceeded the reference QTc values. The lowest automated QTc values were 445 and 447 ms for first-, 463 and 451 ms for seventh-grade male and female students, respectively.</p><p>Conclusion: The reference values for the automatic screening of QT prolongation were 445 and 450 ms in first- and seventh-grade students, respectively.</p>
Journal
-
- Pediatric Cardiology and Cardiac Surgery
-
Pediatric Cardiology and Cardiac Surgery 37 (1), 29-34, 2021-04-01
Japanese Society of Pediatric Cardiology and Cardiac Surgery