書誌事項
- タイトル別名
-
- Results of the acceptance test on Tandetron II in DMRC
この論文をさがす
抄録
Tandetron II in DMRC (Model-4130 AMS, HVEE) is an accelerator mass spectrometer for the ^<14>C dating. The tuning up had been repeated many times after the installation. Finally, the "Acceptance Test", the test to estimate the measurement accuracy of spectrometer itself, was performed on 22 and 23 January, 1999. The results of the test and the accuracy specification are shown in Table 2. The measurement accuracies for the ^<14>C/^<12>C and ^<13>C/^<12>C ratio is quite good and well within the specification. Tandetron II proved it's high accuracy, high statistics and high reproducability of the measurement. It is also very stable during long time measurement. Tandetron II is now at the starting point of the measurement. It should need, however, further checking and tuning up by measuring the known samples. The routine measurement is expected to start from the summer in 1999.
収録刊行物
-
- 名古屋大学加速器質量分析計業績報告書
-
名古屋大学加速器質量分析計業績報告書 10 18-23, 1999-03
名古屋大学年代測定資料研究センター 天然放射性元素測定小委員会