Determination of Electrical Parameters for Skin during Galvanic Skin Reflex from Continuous Measurement
この論文をさがす
抄録
Skin impedance satisfies the Cole-Cole arc's law. The change of skin impedance during GSR (Galvanic Skin Reflex) can be expressed by the change of equivalent parallel resistance approximately. Using these characteristics, the complicated change of skin impedance during GSR can be determined continuously from the measurement value in one frequency point.
収録刊行物
-
- Memoirs of the Faculty of Engineering, Okayama University
-
Memoirs of the Faculty of Engineering, Okayama University 23 (1), 61-68, 1988-11-22
Faculty of Engineering, Okayama University
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390853649478302208
-
- NII論文ID
- 120002307322
-
- NII書誌ID
- AA10699856
-
- ISSN
- 04750071
-
- DOI
- 10.18926/15467
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- IRDB
- CiNii Articles