Accuracy of the ClearSight system in patients undergoing abdominal aortic aneurysm surgery
書誌事項
- 公開日
- 2019
- 資源種別
- journal article
- DOI
-
- 10.1007/s00540-019-02632-6
- 公開者
- Tokyo : Springer Nature Japan
この論文をさがす
説明
The ClearSight™ device monitors continuous pressure and cardiac output via pulse contour analysis. ClearSight™, however, may not be reliable in patients with reduced peripheral perfusion caused by high peripheral resistance. This study aimed to elucidate the accuracy and trending ability of ClearSight™ in patients undergoing abdominal aortic aneurysm (AAA) surgery by comparing the ClearSightThe study included 20 patients who underwent elective AAA surgery. CIBland-Altman analysis showed a wide limit of agreement between CIClearSight™ was not sufficiently accurate in patients undergoing AAA surgery. The tracking ability of ClearSight™ after aortic clamping was below the acceptable limit.
収録刊行物
-
- Journal of anesthesia / Japanese Society of Anesthesiologists
-
Journal of anesthesia / Japanese Society of Anesthesiologists 33 (3), 364-371, 2019
Tokyo : Springer Nature Japan
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1521699230092581120
-
- NII論文ID
- 40021917665
-
- NII書誌ID
- AA10852931
-
- ISSN
- 09138668
- 14388359
-
- NDL書誌ID
- 029742678
-
- PubMed
- 30904953
-
- 本文言語コード
- en
-
- 資料種別
- journal article
-
- NDL 雑誌分類
-
- ZS35(科学技術--医学--外科学・整形外科学・麻酔学)
-
- データソース種別
-
- NDLサーチ
- Crossref
- CiNii Articles
- KAKEN
- OpenAIRE