Study on degradation of VUV-sensitivity of MPPC for liquid xenon scintillation detector by radiation damage in MEG II experiment

Search this article

Description

In the MEG II experiment, the liquid xenon gamma-ray detector uses Multi-Pixel Photon Counters (MPPC) which are sensitive to vacuum ultraviolet (VUV) light under a high-intensity muon beam environment. In the commissioning phase of the detector with the beam, a significant degradation in the photon detection efficiency (PDE) for VUV light was found, while the degradation in the PDE for visible light was much less significant. This implies that the radiation damage is localized to the surface of the MPPC where incoming VUV photons create electron-hole pairs. It was also found that the PDE can recover to the original level by thermal annealing.

17 pages, 11 figures

Journal

Citations (1)*help

See more

References(9)*help

See more

Related Projects

See more

Report a problem