Bibliographic Information
- Other Title
-
- トウ イン ニ オケル カプセル ナイシキョウ ケンサ ノ ガゾウ フリョウ ジレイ ト タイサク ニ ツイテ
- Assessment of accidental events in images during capsule endoscopy
- カプセル内視鏡における画像不良
Search this article
Description
Recently, there have been some reports of electromagnetic interference during capsule endoscopy(CE)in regards to other medical devices and facility environment. These accidental events have not been analyzed sufficiently although academic societies and related ministries are proposing guidelines with cautious attention. Based on this, detailed analyses and development of solution are required. In this study, we review our experience with cases of accidental events during CE and assessed the causes and preventive solutions. A total of149CE cases(male :89, female :60, mean age :56±18.8)from January 2017 until December 2019 were analyzed retrospectively. Four cases(2%)with defective images were noted. Detailed events were as follows ;1 disorder of the sensor array, 2 recorder freezes, and 1 electromagnetic interference. These mechanical and electromagnetic troubles should be assessed and shared among medical staffs and manufacturers to propose possible solutions and perform CE safely.