Cerium X-ray Spectra without Filtering and their Application to High-contrast Angiography

この論文をさがす

抄録

type:P(論文)

The cerium-target x-ray tube is useful in order to perform cone-beam K-edge angiography because K-series characteristic x rays from the cerium target are absorbed effectively by iodine-based contrast media. The x-ray generator consists of a main controller, an x-ray tube unit with a high-voltage circuit and an insulation transformer, and a personal computer. The tube is a glass-enclosed diode with a cerium target and a 0.5-mm-thick beryllium window. The maximum tube voltage and current were 65 kV and 0.4 mA, respectively, and the focal-spot sizes were 1.2×0.8 mm. Sharp cerium K-series characteristic x rays were observed without using a filter, and the x-ray intensity was 209μGy/s at 1.0 m from the source with a tube voltage of 60 kV and a current of 0.40 mA. Angiography was performed with a computed radiography system using iodine-based microspheres 15 μm in diameter. In angiography of non-living animals, we observed fine blood vessels of approximately 100 μm with high contrasts.