光ファイバー温度計の計測精度に関する研究(第1報,火炎温度計測センサーの特性)

書誌事項

タイトル別名
  • ヒカリ ファイバー オンドケイ ノ ケイソク セイド ニ カンスル ケンキュウ
  • Studies on Measurement Accuracy of the Optical Fiber Thermometer(1st Report;Characteristics of Sapphire Rod Sensor for Flame Temperature Measurement)

この論文をさがす

抄録

In order to measure accurately the flame temperature by means of the optical fiber thermometer, the fundamental sensor characteristics were made clear in the following points by using two kinds of calibration apparatus with the blackbody furnace and the semiconductor laser light source. The view angles of the flat-edge type sensor and the beveled-edge type sensor were very narrow within ±10° in pitch angle while the yaw angle characteristic of the beveled one was within ±15°. The decrement of radiation intensity due to the sapphire rod sensor and the optical fiber cable, which dominates the measurement accuracy of the flame temperature, was also calibrated quantitatively by using the blackbody furnace, and it was so large as 58% in the present beveled-edge sensor.

長崎大学工学部研究報告, 24(43), pp.129-134; 1994