DETERMINATION OF FIXED POINTS IN THE LOW TEMPERATURE WITH A HYDROGEN THERMOMETER

  • Shin’ichi Aoyama
    The Cryogenic Laboratory, Research Institute for Iron, Steel, and Other Metals, Imperial University
  • Eizô Kanda
    The Cryogenic Laboratory, Research Institute for Iron, Steel, and Other Metals, Imperial University

書誌事項

公開日
1935-10-01
権利情報
  • https://academic.oup.com/pages/standard-publication-reuse-rights
DOI
  • 10.1246/bcsj.10.472
公開者
Oxford University Press (OUP)

この論文をさがす

説明

<jats:title>Abstract</jats:title> <jats:p>A hydrogen gas thermometer for the standard of the low temperature measurement was constructed wholly of glass. The accuracy of temperature obtained, ±0.036° at 0°C. and ±0.025° at −200°C., was discussed, the volume of thermometer bulb, the dead spaces, and the temperature of the room being taken into consideration. The determination of the boiling points of liquid nitrogen and liquid oxygen and that of the sublimation point of solid carbon dioxide were made with this thermometer. The calibration of copper-constantan thermocouples was also made against the thermometer at various low temperatures.</jats:p>