- Shin’ichi Aoyama
- The Cryogenic Laboratory, Research Institute for Iron, Steel, and Other Metals, Imperial University
-
- Eizô Kanda
- The Cryogenic Laboratory, Research Institute for Iron, Steel, and Other Metals, Imperial University
書誌事項
- 公開日
- 1935-10-01
- 権利情報
-
- https://academic.oup.com/pages/standard-publication-reuse-rights
- DOI
-
- 10.1246/bcsj.10.472
- 公開者
- Oxford University Press (OUP)
この論文をさがす
説明
<jats:title>Abstract</jats:title> <jats:p>A hydrogen gas thermometer for the standard of the low temperature measurement was constructed wholly of glass. The accuracy of temperature obtained, ±0.036° at 0°C. and ±0.025° at −200°C., was discussed, the volume of thermometer bulb, the dead spaces, and the temperature of the room being taken into consideration. The determination of the boiling points of liquid nitrogen and liquid oxygen and that of the sublimation point of solid carbon dioxide were made with this thermometer. The calibration of copper-constantan thermocouples was also made against the thermometer at various low temperatures.</jats:p>
収録刊行物
-
- Bulletin of the Chemical Society of Japan
-
Bulletin of the Chemical Society of Japan 10 (10), 472-481, 1935-10-01
Oxford University Press (OUP)