A standardized evaluation of artefacts from metallic compounds during fast MR imaging

書誌事項

公開日
2016-10
資源種別
journal article
DOI
  • 10.1259/dmfr.20160094
公開者
Oxford University Press (OUP)

この論文をさがす

説明

Metallic compounds present in the oral and maxillofacial regions (OMRs) cause large artefacts during MR scanning. We quantitatively assessed these artefacts embedded within a phantom according to standards set by the American Society for Testing and Materials (ASTM).Seven metallic dental materials (each of which was a 10-mmTukey-Kramer post hoc tests were used for statistical comparisons. For most materials, scanning sequences eliciting artefact volumes in the following (ascending) order FSE-TThe availability of standardized information on artefact size and configuration during MRI will enhance diagnosis when faced with metallic compounds in the OMR.