Complex impedance measurements of calorimeters and bolometers: Correction for stray impedances
-
- Mark. A. Lindeman
- University of Wisconsin , Madison, Wisconsin 53706
-
- Kathleen A. Barger
- University of Wisconsin , Madison, Wisconsin 53706
-
- Donald E. Brandl
- University of Wisconsin , Madison, Wisconsin 53706
-
- S. Gwynne Crowder
- University of Wisconsin , Madison, Wisconsin 53706
-
- Lindsay Rocks
- University of Wisconsin , Madison, Wisconsin 53706
-
- Dan McCammon
- University of Wisconsin , Madison, Wisconsin 53706
説明
<jats:p>Impedance measurements provide a useful probe of the physics of bolometers and calorimeters. We describe a method for measuring the complex impedance of these devices. In previous work, stray impedances and readout electronics of the measurement apparatus have resulted in artifacts in the impedance data. The technique allows experimenters to find an independent Thevenin or Norton equivalent circuit for each frequency. This method allows experimenters to easily isolate the device impedance from the effects of parasitic impedances and frequency dependent gains in amplifiers.</jats:p>
収録刊行物
-
- Review of Scientific Instruments
-
Review of Scientific Instruments 78 (4), 2007-04-01