Waveform Analysis of a 240-Pixel TES Array for X-Rays and Charged Particles Using a Function of Triggering Neighboring Pixels

書誌事項

公開日
2020-04-13
資源種別
journal article
権利情報
  • https://www.springer.com/tdm
  • https://www.springer.com/tdm
DOI
  • 10.1007/s10909-020-02449-8
公開者
Springer Science and Business Media LLC

この論文をさがす

説明

A useful function, called group trigger, is implemented in a data acquisition system of a 240-pixel X-ray transition-edge sensor array to store simultaneous waveforms from any desired set of pixels. It works as a diagnostic tool which can record both primary pulses and associated crosstalk events with enough freedom to optimize the data output. Under a high rate of charged particle background, such as in an accelerator, investigating signals from any combination of trigger patterns for a particular event is of high importance. We utilized this function throughout an entire experiment at J-PARC, measuring kaonic atom X-rays in 2018. This was the first time a group trigger was applied to exotic-atom spectroscopy. In this experiment, the primary trigger was distributed to the four physical nearest neighbor pixels for practical purposes. The function clarified the effect of thermal and electrical cross talk from X-rays and charged particles, resulting in a better understanding of the behavior of the TES array in such a high-background environment.

収録刊行物

被引用文献 (3)*注記

もっと見る

参考文献 (6)*注記

もっと見る

関連プロジェクト

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ