Effect of endoscope flexibility on tissue dissection profile assessed with pulsed water jet device: ensuring safety, efficacy, and handling of thin devices for neuroendoscopic surgery

DOI DOI PDF Web Site Web Site ほか2件をすべて表示 一部だけ表示 被引用文献1件 参考文献12件 オープンアクセス

説明

<jats:title>Abstract</jats:title><jats:sec> <jats:title>Objective</jats:title> <jats:p>We developed an actuator-driven pulsed water jet device (ADPJ) for flexible neuroendoscopy to achieve effective tissue dissection with vasculature preservation. Although flexibility is a strong advantage for minimally invasiveness, the effect of the ductile curvature on the dissection profiles remains unknown. The purpose of this study was to clarify the impact of the curvature change of the ADPJ connecting tube on the dissection safety and efficacy.</jats:p> </jats:sec><jats:sec> <jats:title>Results</jats:title> <jats:p>Three ADPJ connecting tubes with different inner diameters (1.0, 0.75, 0.5 mm) were used to dissect the brain phantom. They were bent at 3 angles: 0°, 60°, and 120°. The dissection profiles were evaluated using the mean depth and coefficient of variation (CV) for efficacy and safety, respectively.The larger inner diameter connecting tube dissected more deeply. The dissection depth was not changed regardless of the curvature degree in each tube. There was no significant difference in CVs regardless of inner diameter and curvature. The ductile curvature of the flexible neuroendoscope did not affect the efficacy and safety of the ADPJ dissection profile. Among the numerous instruments, tube-formed devices, including suction and injecting devices such as ADPJ, can be used safely and effectively without flexibility-related limitations.</jats:p> </jats:sec>

収録刊行物

  • BMC Research Notes

    BMC Research Notes 14 (1), 64-, 2021-02-17

    Springer Science and Business Media LLC

被引用文献 (1)*注記

もっと見る

参考文献 (12)*注記

もっと見る

関連プロジェクト

もっと見る

問題の指摘

ページトップへ