Eye lens dose for medical staff assisting patients during computed tomography: comparison of several types of radioprotective glasses

抄録

<jats:title>Abstract</jats:title> <jats:p>Medical staff sometimes assists patients in the examination room during computed tomography (CT) scans for several purposes. This study aimed to investigate the dose reduction effects of four radioprotective glasses with different lead equivalents and lens shapes. A medical staff phantom was positioned assuming body movement restraint of the patient during chest CT, and H<jats:sub>p</jats:sub>(3) at the eye surfaces of the medical staff phantom and inside the lens of the four types of radioprotective glasses were measured by changing the distance of the staff phantom from the gantry, eye height, and width of the nose pad. The H<jats:sub>p</jats:sub>(3) at the right eye surface with glasses of 0.50–0.75 mmPb and 0.07 mmPb was approximately 83.5% and 58.0%, respectively, lower than that without radioprotective glasses. The dose reduction rates at left eye surface increased with over-glass type glasses by 14%–28% by increasing the distance from the CT gantry to the staff phantom from 25 to 65 cm. The dose reduction rates at the left eye surface decreased with over-glass type glasses by 26%–31% by increasing the height of the eye lens for the medical staff phantom from 130 to 170 cm. The H<jats:sub>p</jats:sub>(3) on the left eye surface decreased by 46.9% with the widest nose pad width compared to the narrowest nose pad width for the glasses with adjustable nose pad width. The radioprotective glasses for staff assisting patients during CT examinations should have a high lead equivalent and no gap around the nose and under the front lens.</jats:p>

収録刊行物

参考文献 (20)*注記

もっと見る

関連プロジェクト

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ