- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Knowledge Graph Search feature is available on CiNii Labs
- Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
A New Method for Monitoring Deep Body Temperature from the Skin Surface
-
- R. H. Fox
- National Institute for Medical Research (Hampstead Laboratories), Holly Hill, London
-
- A. J. Solman
- Clinical Research Centre, Watford Road, Harrow, Middlesex
-
- R. Isaacs
- Clinical Research Centre, Watford Road, Harrow, Middlesex
-
- A. J. Fry
- National Institute for Medical Research (Hampstead Laboratories), Holly Hill, London
-
- I. C. MacDonald
- National Institute for Medical Research (Hampstead Laboratories), Holly Hill, London
Search this article
Description
<jats:p>1. A new technique for monitoring the deep body temperature is described. The technique depends on creating a zone of zero heat-flow across the body shell; this brings the deep body temperature to the skin surface where it is measured with a simple electronic thermometer.</jats:p> <jats:p>2. The new device gives a temperature closely comparable with other methods for measuring the deep body temperature in the resting subject, and is simple to use and socially acceptable.</jats:p>
Journal
-
- Clinical Science
-
Clinical Science 44 (1), 81-86, 1973-01-01
Portland Press Ltd.
- Tweet
Details 詳細情報について
-
- CRID
- 1360855569288852480
-
- ISSN
- 03010538
-
- Data Source
-
- Crossref