Assessing coronary blood flow dynamics with the TIMI frame count method: Comparison with simultaneous intracoronary Doppler and ultrasound

この論文をさがす

説明

<jats:title>Abstract</jats:title><jats:p>This study compared the TIMI frame count (TFC), which has been proposed as a method for quantifying coronary blood flow, with coronary flow and microvascular function measured with intracoronary Doppler and intracoronary ultrasound. Coronary blood flow volume was calculated from coronary blood velocity (by intracoronary Doppler) and lumen area (by intracoronary ultrasound) in the LAD in 46 post–heart transplant patients at baseline and after intracoronary adenosine. TFC correlated significantly with average peak coronary blood velocity (r = −0.42; <jats:italic>P</jats:italic> = 0.004) and coronary lumen area (r = 0.39; <jats:italic>P</jats:italic> = 0.008), but not with coronary blood flow volume (r = −0.01; <jats:italic>P</jats:italic> = 0.96) or the coronary flow reserve response to adenosine (r = 0.09; <jats:italic>P</jats:italic> = 0.58). In conclusion, TFC is a simple method of assessing coronary blood velocity but not volumetric flow. While TFC does not predict coronary flow reserve, as a measure of velocity it does provide an assessment of basal microvascular tone, information that is complementary to that afforded by flow reserve measurements. Cathet Cardiovasc Intervent 2001;53:459–463. © 2001 Wiley‐Liss, Inc.</jats:p>