Spectral and sequence analysis of finger blood pressure variability. Comparison with analysis of intra-arterial recordings.

  • S Omboni
    Cattedra di Medicina Interna, Ospedale S. Gerardo, Monza, Italy.
  • G Parati
    Cattedra di Medicina Interna, Ospedale S. Gerardo, Monza, Italy.
  • A Frattola
    Cattedra di Medicina Interna, Ospedale S. Gerardo, Monza, Italy.
  • E Mutti
    Cattedra di Medicina Interna, Ospedale S. Gerardo, Monza, Italy.
  • M Di Rienzo
    Cattedra di Medicina Interna, Ospedale S. Gerardo, Monza, Italy.
  • P Castiglioni
    Cattedra di Medicina Interna, Ospedale S. Gerardo, Monza, Italy.
  • G Mancia
    Cattedra di Medicina Interna, Ospedale S. Gerardo, Monza, Italy.

書誌事項

公開日
1993-07
DOI
  • 10.1161/01.hyp.22.1.26
公開者
Ovid Technologies (Wolters Kluwer Health)

この論文をさがす

説明

<jats:p>The aim of our study was to assess whether the Finapres device is able to accurately monitor not only average blood pressure values but also blood pressure variability. To examine this issue, we analyzed 30-minute recordings of finger and intra-arterial pressure simultaneously obtained at rest in 14 patients. We compared systolic blood pressure, diastolic blood pressure, mean arterial pressure, pulse interval (the reciprocal of heart rate), overall variability (standard deviation), and specific time-domain and frequency-domain components. Systolic blood pressure, diastolic blood pressure, mean arterial pressure, and pulse interval spectral powers were computed by fast Fourier transform over three frequency bands: low frequency (0.025 to 0.07 Hz), midfrequency (0.07 to 0.14 Hz), and high frequency (0.14 to 0.35 Hz). The coherence, ie, the degree of association between blood pressure and pulse interval powers obtained by the two techniques, was also assessed. Standard deviations of diastolic blood pressure, mean arterial pressure, and pulse interval were similar when assessed from the two recordings, whereas standard deviation of systolic blood pressure was overestimated by analysis of finger pressure recordings. All powers of diastolic blood pressure and mean arterial pressure and high-frequency powers of systolic blood pressure estimated from analysis of finger blood pressure tracings were superimposable to those obtained by analyzing invasive recordings. Low-frequency and midfrequency powers of intra-arterial systolic blood pressure were significantly overestimated by the analysis of finger blood pressure tracings (+13.7 +/- 4.4 mm Hg2, P < .01, and +2.3 +/- 0.9 mm Hg2, P < .05).(ABSTRACT TRUNCATED AT 250 WORDS)</jats:p>

収録刊行物

  • Hypertension

    Hypertension 22 (1), 26-33, 1993-07