- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Knowledge Graph Search feature is available on CiNii Labs
- 【Updated on June 30, 2025】Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
Single-isocenter volumetric-modulated Dynamic WaveArc therapy for two brain metastases
Search this article
Description
A new irradiation technique, volumetric-modulated Dynamic WaveArc therapy (VMDWAT), based on sequential non-coplanar trajectories, can be performed using the Vero4DRT. This planning study compared the dose distribution and treatment time between single-isocenter volumetric-modulated arc therapy (VMAT) with multiple straight non-coplanar arcs and single-isocenter VMDWAT in patients with two brain metastases.Twenty patients with two planning target volumes exceeding 2.0 cmVMDWAT significantly improved both mRTOG-CI and mIP-CI and reduced the volume of normal brain tissue receiving 25 and 28 Gy compared to VMAT. The two modalities did not significantly differ in terms of the volume of normal brain tissue receiving 5, 10, 12, 15, and 20 Gy. The mean treatment time was significantly shorter in the VMDWAT group.VMDWAT significantly improved dose distribution in a shorter treatment time compared to VMAT in patients treated for two brain metastases. Single-isocenter VMDWAT may thus be a promising treatment for two brain metastases.
Journal
-
- Japanese Journal of Radiology
-
Japanese Journal of Radiology 37 (8), 619-625, 2019-06-22
Springer Science and Business Media LLC