- G. Fiksel
- University of Rochester Laboratory for Laser Energetics, , 250 East River Rd., Rochester, New York 14623-1299, USA
-
- F. J. Marshall
- University of Rochester Laboratory for Laser Energetics, , 250 East River Rd., Rochester, New York 14623-1299, USA
-
- C. Mileham
- University of Rochester Laboratory for Laser Energetics, , 250 East River Rd., Rochester, New York 14623-1299, USA
-
- C. Stoeckl
- University of Rochester Laboratory for Laser Energetics, , 250 East River Rd., Rochester, New York 14623-1299, USA
抄録
<jats:p>The spatial resolution of two types of imaging plates, Fuji BAS-TR and Fuji BAS-SR, has been measured using a knife-edge x-ray source of 8-keV Cu Kα radiation. The values for the spatial resolution, defined as the distance between 10% and 90% levels of the edge spread function, are 94 μm and 109 μm, respectively. The resolution values are important for quantitative analysis of x-ray and particle imaging and spectroscopic diagnostics.</jats:p>
収録刊行物
-
- Review of Scientific Instruments
-
Review of Scientific Instruments 83 (8), 086103-, 2012-08-01
AIP Publishing
- Tweet