<jats:p>The spatial resolution of two types of imaging plates, Fuji BAS-TR and Fuji BAS-SR, has been measured using a knife-edge x-ray source of 8-keV Cu Kα radiation. The values for the spatial resolution, defined as the distance between 10% and 90% levels of the edge spread function, are 94 μm and 109 μm, respectively. The resolution values are important for quantitative analysis of x-ray and particle imaging and spectroscopic diagnostics.</jats:p>