Note: Spatial resolution of Fuji BAS-TR and BAS-SR imaging plates

  • G. Fiksel
    University of Rochester Laboratory for Laser Energetics, , 250 East River Rd., Rochester, New York 14623-1299, USA
  • F. J. Marshall
    University of Rochester Laboratory for Laser Energetics, , 250 East River Rd., Rochester, New York 14623-1299, USA
  • C. Mileham
    University of Rochester Laboratory for Laser Energetics, , 250 East River Rd., Rochester, New York 14623-1299, USA
  • C. Stoeckl
    University of Rochester Laboratory for Laser Energetics, , 250 East River Rd., Rochester, New York 14623-1299, USA

抄録

<jats:p>The spatial resolution of two types of imaging plates, Fuji BAS-TR and Fuji BAS-SR, has been measured using a knife-edge x-ray source of 8-keV Cu Kα radiation. The values for the spatial resolution, defined as the distance between 10% and 90% levels of the edge spread function, are 94 μm and 109 μm, respectively. The resolution values are important for quantitative analysis of x-ray and particle imaging and spectroscopic diagnostics.</jats:p>