Reflectron time-of-flight mass spectrometer for laser photodissociation

  • D. S. Cornett
    Department of Chemistry, University of Georgia, Athens, Georgia 30602
  • M. Peschke
    Department of Chemistry, University of Georgia, Athens, Georgia 30602
  • K. LaiHing
    Department of Chemistry, University of Georgia, Athens, Georgia 30602
  • P. Y. Cheng
    Department of Chemistry, University of Georgia, Athens, Georgia 30602
  • K. F. Willey
    Department of Chemistry, University of Georgia, Athens, Georgia 30602
  • M. A. Duncan
    Department of Chemistry, University of Georgia, Athens, Georgia 30602

書誌事項

公開日
1992-04-01
DOI
  • 10.1063/1.1143135
公開者
AIP Publishing

この論文をさがす

説明

<jats:p>We describe a new reflectron time-of-flight mass spectrometer configuration for laser photodissociation of mass-selected ions and the initial performance characteristics observed for this instrument. Ions are produced by laser photoionization within the acceleration region of the instrument or by laser vaporization in an external pulsed-nozzle cluster ion source. Mass selection is accomplished with pulsed deflection plates at the end of an initial drift section. Laser photodissociation of selected ions takes place at the turning point in the ion trajectory in the reflectron. The transit time through a second drift section defines the fragment ion masses. Optimized operating conditions and the role of mass discrimination in this instrument are discussed.</jats:p>