A dramatic increase in the positive blood culture rates of Helicobacter cinaedi: the evidence of differential detection abilities between the Bactec and BacT/Alert systems
書誌事項
- 公開日
- 2015-11
- 権利情報
-
- https://www.elsevier.com/tdm/userlicense/1.0/
- DOI
-
- 10.1016/j.diagmicrobio.2015.07.017
- 公開者
- Elsevier BV
この論文をさがす
説明
In our hospital, positive blood culture rates of Helicobacter cinaedi dramatically increased after introducing the Bactec system. A simulated culture model of H. cinaedi bacteremia demonstrated no positive signals using the BacT/Alert system, despite efficient growth in bottles. Clinically suspected H. cinaedi bacteremia should be monitored more closely when using the BacT/Alert system, preferably with subcultivation after 7days of incubation.
収録刊行物
-
- Diagnostic Microbiology and Infectious Disease
-
Diagnostic Microbiology and Infectious Disease 83 (3), 232-233, 2015-11