Ultrashort laser based two-photon phase-resolved fluorescence lifetime measurement method

説明

This paper presents a two-photon phase-resolved fluorescence-lifetime measurement method based on the use of an ultrashort pulse laser. The proposed method also involves the use of a lock-in amplifier to control the phase difference between the reference and fluorescence signals, thereby facilitating the use of an alternative method for determining fluorescence lifetimes. Verification of the fluorescence lifetimes as measured in this study was performed using rhodamine B and a cellular thermoprobe as samples. In this study, we assume that the fluorescence decay was monoexponential in all cases. Rhodamine B was observed to exhibit an average fluorescence lifetime of 2.15 ns, whereas a temperature sensitivity of 1.39 ns C