Clinical evaluation of the Sorin Xtra® autotransfusion system

  • EP Overdevest
    Department of Extracorporeal Circulation and Blood Management, Catharina Hospital Eindhoven, Eindhoven, the Netherlands
  • PWJ Lanen
    Department of Extracorporeal Circulation and Blood Management, Catharina Hospital Eindhoven, Eindhoven, the Netherlands
  • JCM Feron
    Department of Extracorporeal Circulation and Blood Management, Catharina Hospital Eindhoven, Eindhoven, the Netherlands
  • JWH van Hees
    Department of Extracorporeal Circulation and Blood Management, Catharina Hospital Eindhoven, Eindhoven, the Netherlands
  • MESH Tan
    Department of Cardiothoracic Surgery, Catharina Hospital Eindhoven, Eindhoven, the Netherlands

この論文をさがす

説明

<jats:p> The performance of the Sorin Xtra® Autotransfusion System (ATS) was studied in 62 patients undergoing coronary artery bypass grafting. Blood was collected intraoperatively and washed using three different wash sets in 4 groups. Both collected and washed blood were analysed for hemoglobin levels and hematocrit, concentrations of proteins, albumin, heparin and plasma free hemoglobin (PFH) were determined, erythrocytes, platelets and leukocytes were counted. </jats:p><jats:p> Hematocrit measurements of the Xtra® were compared with laboratory measurements to study the accuracy of the Xtra® hematocrit sensor. In addition, the red blood cell recovery rate and elimination rates were calculated to evaluate the clinical performance of the Xtra®. </jats:p><jats:p> The Xtra® ATS produced a volume of concentrated red blood cells with an average hematocrit from 58% to 63%, depending on the size of the bowl and the chosen default program. In all bowl sizes and programs, the Xtra® Hct-out measurement underestimated the CELL-DYN measurement by approximately 15%. </jats:p><jats:p> The calculated recovery rates for red blood cells (RBC) in the 4 groups ranged from 86.7% to 91.6%. Elimination rates were calculated in each group for proteins (96.8-99.2%), albumin (96.4-98.7%), plasma free hemoglobin (83.6–91.2%), heparin (98.8-99.9%), platelets (82.4-94.3%) and white blood cells (28.6-42.3%). </jats:p><jats:p> The Xtra® ATS can be appealing for its performance by producing high hematocrit levels in the washed RBC volume, while keeping RBC recovery rate at the same high level (≈ 90%) as in its predecessor, the Electa® Autotransfusion System. </jats:p>

収録刊行物

  • Perfusion

    Perfusion 27 (4), 278-283, 2012-03-29

    SAGE Publications