Technical Note: Comparison between single and multiview simulated DXA configurations for reconstructing the 3D shape and bone mineral density distribution of the proximal femur

抄録

<jats:sec><jats:title>Purpose:</jats:title><jats:p>Dual‐energy x‐ray absorptiometry (DXA) is used in clinical routine to provide a two‐dimensional (2D) analysis of the bone mineral density (BMD). 3D reconstruction methods from 2D DXA images could improve the BMD analysis. To find the optimal configuration that should be used in clinical routine, this paper relies on a 3D reconstruction method from DXA images to compare the accuracy that can be obtained from one single‐view and from multiview DXA images (two to four projections).</jats:p></jats:sec><jats:sec><jats:title>Methods:</jats:title><jats:p>The 3D reconstruction method uses a statistical model and a nonrigid registration technique to recover in 3D the shape and the BMD distribution of the proximal femur. The accuracy was evaluated<jats:italic>in vivo</jats:italic> by comparing 3D reconstructions obtained from simulated DXA images of 30 patients (using between one and four DXA views) with quantitative computed tomography reconstructions.</jats:p></jats:sec><jats:sec><jats:title>Results:</jats:title><jats:p>This comparison showed that the use of one single DXA provides accurate 3D reconstructions (mean shape accuracy of 1.0 mm and BMD distribution errors of 7.0%). Among the multiview configurations, the use of two views (0° and 45°) was the best compromise, increasing the accuracy of pose (mean accuracy of 0.7°/1.2°/0.9° against 1.0°/3.5°/3.3° for the single view), reducing slightly the BMD errors (5.7%) while maintaining the same shape accuracy.</jats:p></jats:sec><jats:sec><jats:title>Conclusions:</jats:title><jats:p>The use of two views constitutes an interesting configuration when multiview DXA devices are available in clinical routine. However, the use of only one single view remains an accurate solution to recover the shape and the BMD distribution in 3D, with the advantage of a higher potential for clinical translation.</jats:p></jats:sec>

収録刊行物

被引用文献 (1)*注記

もっと見る

キーワード

問題の指摘

ページトップへ