Three-dimensional Stress Distribution Around a Dental Implant at Different Stages of Interface Development

  • L. Borchers
    Zentrum Zahn-, Mund-, und Kieferheilkunde der Medizinischen Hochschule Hannover, Karl-Wiechert-Allee 9, D-3000 Hannover 61, Federal Republic of Germany
  • P. Reichart
    Zentrum Zahn-, Mund-, und Kieferheilkunde der Medizinischen Hochschule Hannover, Karl-Wiechert-Allee 9, D-3000 Hannover 61, Federal Republic of Germany

書誌事項

公開日
1983-02
権利情報
  • https://journals.sagepub.com/page/policies/text-and-data-mining-license
DOI
  • 10.1177/00220345830620021401
公開者
SAGE Publications

この論文をさがす

説明

<jats:p>An anchor-type ceramic dental implant contained in a section of the mandibular molar region was modeled for finite element stress analysis. The distributions of stresses in the bone around the implant due to axial and transverse loading were calculated for different stages of normal and pathological development of the implant-bone interface. Highest stress concentrations are observed in the crestal region with transverse loading when spongy bone surrounds the implant immediately after surgery. Development of a lamina dura around the implant slightly reduces severe stress peaks, whereas ingrowth of connective tissue almost eliminates them.</jats:p>