- V. L. Sailor
- Brookhaven National Laboratory, Upton, New York
-
- H. L. Foote
- Brookhaven National Laboratory, Upton, New York
-
- H. H. Landon
- Brookhaven National Laboratory, Upton, New York
-
- R. E. Wood
- Brookhaven National Laboratory, Upton, New York
書誌事項
- 公開日
- 1956-01-01
- DOI
-
- 10.1063/1.1715358
- 公開者
- AIP Publishing
この論文をさがす
説明
<jats:p>A crystal spectrometer for neutron measurements has been developed which has relatively good resolution over an effective energy range from 0.08 to >20 ev. The figure of merit which has been obtained is 0.171 μsec/m with Be (123̄1) and 0.127 μsec/m with Be (224̄2) monochromator. The good resolution is made possible by a multiple channel collimator (Soller slit) of novel design. The collimator is described in detail and the resolution function for the system is derived and discussed. It is shown that the resolution of the system is essentially independent of the mosaic spread of the crystal. Experimental results are presented which illustrate the performance of the instrument.</jats:p>
収録刊行物
-
- Review of Scientific Instruments
-
Review of Scientific Instruments 27 (1), 26-34, 1956-01-01
AIP Publishing