Microbore high-performance liquid chromatography and chemiluminescence detection of Dns-amino acids
書誌事項
- 公開日
- 1984-01
- 権利情報
-
- https://www.elsevier.com/tdm/userlicense/1.0/
- DOI
-
- 10.1016/s0021-9673(00)96178-8
- 公開者
- Elsevier BV
この論文をさがす
説明
Abstract Application of the chemiluminescence (CL) post-column reaction to microbore column (250 × 1 mm I.D., 10 μm) high-performance liquid chromatography is described for the sensitive detection of Dns-amino acids. Dns-alanine, -valine, -isoleucine and -phenylalanine, separated on a reversed-phase column, were detected with CL, generated by the reaction with bis(2,4,6-trichlorophenyl) oxalate and hydrogen peroxide. The detection limits are of the order of 2 × 10 -16 mol (S/N 3) and the average deviation for the peak height of 10 fmol of each Dns-amino acid was 1.3% ( n 5).
収録刊行物
-
- Journal of Chromatography A
-
Journal of Chromatography A 316 501-505, 1984-01
Elsevier BV