An accumulative x-ray streak camera with sub-600-fs temporal resolution and 50-fs timing jitter
-
- Jinyuan Liu
- Experimental Facilities Division, Argonne National Laboratory, Argonne, Illinois 60439
-
- Jin Wang
- Experimental Facilities Division, Argonne National Laboratory, Argonne, Illinois 60439
-
- Bing Shan
- Department of Physics and J. R. Macdonald Laboratory, Kansas State University, Manhattan, Kansas 66506
-
- Chun Wang
- Department of Physics and J. R. Macdonald Laboratory, Kansas State University, Manhattan, Kansas 66506
-
- Zenghu Chang
- Department of Physics and J. R. Macdonald Laboratory, Kansas State University, Manhattan, Kansas 66506
書誌事項
- 公開日
- 2003-05-19
- DOI
-
- 10.1063/1.1577213
- 公開者
- AIP Publishing
この論文をさがす
説明
<jats:p>We demonstrated that the shot-to-shot timing jitter of an x-ray streak camera was reduced to close to 50 fs when it was triggered by a standard kilohertz laser with 1.2% rms pulse energy fluctuation. This was achieved by improving the response time of deflection plates and the rise time of the ramp pulse generated by a photoconductive switch, and by operating the photoconductive switch at optimum conditions. Furthermore, after reducing the angular distribution of electron bunch, the temporal resolution of the x-ray streak camera operating in accumulation mode was measured to be better than 600 fs.</jats:p>
収録刊行物
-
- Applied Physics Letters
-
Applied Physics Letters 82 (20), 3553-3555, 2003-05-19
AIP Publishing