Examination of the Moxus Modular Metabolic System by the Douglas-bag technique

  • Jon Ingulf Medbø
    National Institute of Occupational Health, Oslo, Norway.
  • Asgeir Mamen
    Sogn og Fjordane University College, Faculty of Teacher Education and Sport, Sogndal, Norway.
  • Fernando G. Beltrami
    Exercise Sciences and Sports Medicine Unit, University of Cape Town, Cape Town, South Africa.

この論文をさがす

説明

<jats:p>The purpose of this study was to examine the performance of the Moxus Modular Metabolic System from AEI Technologies, Inc. using the Douglas-bag method as reference. To achieve this, eight moderately trained subjects cycled for 5 min at constant powers from 50 to 300 W in increments of 50 W. The O<jats:sub>2</jats:sub>uptake was measured simultaneously by both systems during the last minute of each stage. The O<jats:sub>2</jats:sub>uptake reported by the Moxus system was 83 ± 78 mL·min<jats:sup>–1</jats:sup>higher (mean ± SD; ≈3%, +62 µmol·s<jats:sup>–1</jats:sup>, P < 0.001) than that reported by the Douglas-bag method; the bias varied by ≈2% between the subjects. The higher O<jats:sub>2</jats:sub>uptake of the Moxus system was a consequence of 1.4% ± 3.0% higher reported ventilation and 2% ± 3% higher reported O<jats:sub>2</jats:sub>extraction per volume of air breathed. The respiratory exchange ratio (R value) reported by the Moxus system rose proportionally to that of the Douglas-bag method and was 1% ± 2% higher for the range examined (0.75–1.10). Repeated tests of the maximal O<jats:sub>2</jats:sub>uptake showed a variability (coefficient of variation) of 2.5%. The study concluded that measurements by the Moxus system showed some bias and residual variation and, in addition, some systematic differences between the subjects in the O<jats:sub>2</jats:sub>uptake. The R value was reported quite accurately with moderate random error. Although there were some computer software and hardware instability problems that need to be solved, the Moxus system worked quite well and provided data more reliable than those of most commercial instruments.</jats:p>

収録刊行物

被引用文献 (1)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ