Determination of arsenic(III) and arsenic(V) in fresh water by AAS equipped with arsine generation system.
-
- TUJI Haruo
- Konan University
-
- TAMARI Yuzo
- Konan University
-
- KATAGIRI Sayaka
- Konan University
-
- YAMAZAKI Kenji
- Konan University
-
- KUSAKA Yuzuru
- Konan University
Bibliographic Information
- Other Title
-
- 還元気化原子吸光法による陸水中のヒ素(III)及びヒ素(V)の分別定量
- カンゲン キカ ゲンシ キュウコウホウ ニ ヨル リクスイチュウ ノ ヒソ 3
Search this article
Abstract
For the determination of arsenic(III), arsine was first generated from the sample solution at pH 9 with tetrahydroborate, and then it was introduced into a heated quartz glass cell along with argon; and finally the atomic absorption of arsenic was measured. In addition, the total arsenic was determined under acidic condition (5% hydrochloric acid solution) in the same manner. The arsenic(V) was determined by subtracting the value of arsenic(III) from that of total arsenic. This method was applied for ground and river water samples. Detection levels of 2.6-8.0 ppb of arsenic(V) were detected, whereas arsenic(III) was not detected.
Journal
-
- BUNSEKI KAGAKU
-
BUNSEKI KAGAKU 40 (4), T97-T100, 1991
The Japan Society for Analytical Chemistry
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001204051114112
-
- NII Article ID
- 110002906263
-
- NII Book ID
- AN00222633
-
- COI
- 1:CAS:528:DyaK3MXktlGgtrg%3D
-
- NDL BIB ID
- 3713679
-
- ISSN
- 05251931
- http://id.crossref.org/issn/05251931
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- NDL-Digital
- CiNii Articles
-
- Abstract License Flag
- Disallowed