高分子イオン伝導体膜を用いたガルバニ型ガスセンサーによる二酸化窒素のフローインジェクション分析

書誌事項

タイトル別名
  • A galvanic gas detector using polymeric electrolyte film for FIA of nitrogen dioxide.
  • コウブンシ イオン デンドウタイ マク オ モチイタ ガルバニケイ ガス セン

この論文をさがす

抄録

An Au-poly(ethylene oxide)4.5AgCF3SO3-Ag galvanic gas detector, consisting of a disc (13mm in diameter×1.1mm thick) has been developed by utilizing a polymeric electrolyte film (30μm thick). The polymeric films were made by mixing acetonitrile solution with both AgCF3SO3 and polyethylene oxide. The mixture was cast on the disc and the solvent was permitted to evaporate. The working electrode (25nm thick) was made by coating it with vaporized gold. The maximum response appeared within 3 s after the sample (2.6ml) was injected into the nitrogen carrier stream. The flow rate was 40ml min-1. The longest time to return to the base line was 3 min. The maximum current flowing in the external circuit increased linearly relating to the concentration of NO2 in the range 0.28ppm at ambient temperature. Twenty samples could be analyzed per hour. The relative standard deviation of the response for 1.0ppm of NO2 using 2.6ml sample was about 3%. After 200 samples (1ppm of NO2-2.6ml) had been analyzed, the response decreased by about 15% of its original value.

収録刊行物

  • 分析化学

    分析化学 40 (4), T77-T81, 1991

    公益社団法人 日本分析化学会

被引用文献 (1)*注記

もっと見る