Study of a metal-coated hollow-fiber membrane electrode for sugar detection by an amperometric detection method.

Bibliographic Information

Other Title
  • 導電性中空糸膜電極を用いる電気化学検出法による糖検出感度の検討
  • ドウデンセイ チュウクウシ マク デンキョク オ モチイル デンキ カガク ケンシュツホウ ニ ヨル トウ ケンシュツ カンド ノ ケントウ

Search this article

Description

A metal-coated hollow-fiber membrane electrode (MCHE) has been applied as a working electrode of an electrochemical detector in high-performance liquid chromatography. In this paper, the structure of the detector was studied for obtaining a high sensitivity and sharp response on a model of electrochemical oxidation of glucose in an alkaline solution. The MCHE has a tubular structure with an inner volume that broadens the output signal peak with a wide distribution of the specimen concentration due to a residual sodium hydroxide solution. An analysis of the dead volume of the present detector revealed that the volume of holes in the membrane wall was almost equal to the inner volume of the electrode. Only the inside metal layer of the MCHE was made active for electrochemical detection in order to reduce the effect of the dead volume of the membrane holes. The detection limit and the half-width of the peak were down to 0.69 pmol and 1.8 s, respectively.<br>

Journal

  • BUNSEKI KAGAKU

    BUNSEKI KAGAKU 52 (4), 245-251, 2003

    The Japan Society for Analytical Chemistry

References(25)*help

See more

Details 詳細情報について

Report a problem

Back to top