o-ヒドロキシヒドロキノンフタレインと陽イオン性界面活性剤を用いるクロム(VI)の吸光光度定量

書誌事項

タイトル別名
  • キサンテン系誘導体の分析化学への利用 XI  ?Eo?F‐ヒドロキシヒドロキノンフタレインと陽イオン性界面活性剤を用いるクロム(VI)の吸光光度定量
  • Spectrophotometric determination of chromium (VI) with ο-hydroxyhydroquinonephthalein and cationic surface-active agent
  • オミクロン ヒドロキシヒドロキノンフタレイン ト ヨウイオンセイ カイメン カ
  • キサンテン系誘導体の分析化学への利用(第11報)

この論文をさがす

抄録

ο-Hydroxyhydroquinonephthalein (Qn. Ph.)-chromium(VI) {Cr(VI)} complex in the presence of excess cationic surface-active agent (cetyltrimethylammonium chloride) (CTAC) could be used for the spectrophotometric determination of a minute amount of Cr(VI). A maximum absorption wavelength of the complex was 560 nm in a weakly acidic medium, where the apparent molar absorption coefficient was 1.64×105. The Sandell sensitivity was 0.00032μg/cm2 Cr(VI) for an absorbance of 0.001 at the wavelength of 560 nm. The calibration curve was linear in the range 0 to 4μg/10ml Cr(VI). The proposed procedure for the determination of Cr (VI) is as follows. To a 10 ml volumetric flask a sample solution containing less than 4 μg of Cr(VI), 1.0 ml of 2.0×10-2 M CTAC solution and 2.0 ml of 1.0×10-3 M Qn. Ph. methanol solution are added, and the pH of the mixture is adjusted to 5.8 by the addition of Sorensen buffer solution of phosphate. The content is diluted to the mark with water, kept at 60°C for 30 min, and the absorbance of the solution is measured at 560 nm against the reagent blank. The reaction is quite sensitive, and the procedure does not require a long time. This method is applicable to the determination of Cr(VI) in steel in much the same way as previous by reported determination of molybdenum(VI). The mole ratio of Cr(VI), Qn. Ph. and CTAC in the complex was estimated to be 1 : 2 : 2 by the continuous variation and the mole ratio methods.

収録刊行物

  • 分析化学

    分析化学 28 (11), 707-710, 1979

    公益社団法人 日本分析化学会

被引用文献 (5)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ