Ion Orbit Analysis for Quadrupole Mass Spectrometer (II).

  • HIROKI Seiji
    Japan Atomic Energy Research Institute, Naka Fusion Research Establishment
  • ABE Tetsuya
    Japan Atomic Energy Research Institute, Naka Fusion Research Establishment
  • MURAKAMI Yoshio
    Japan Atomic Energy Research Institute, Naka Fusion Research Establishment

Bibliographic Information

Other Title
  • 四極子質量分析計におけるイオンの軌道解析(II)
  • 4キョクシ シツリョウ ブンセキケイ ニ オケル イオン ノ キドウ カイセキ

Search this article

Abstract

Ion orbits in the quadrupole field were calculated for four stability zones using a matrix method. The orbits were compared with those from the approximate formulae previously obtained. The dependence of maximum ion amplitude on initial phase angle was also calculated for x and y directions for respective zones. It has been found that a condition of the second stability zone at the upper tip is useful in obtaining a high resolution peak.

Journal

  • Shinku

    Shinku 36 (1), 1-7, 1993

    The Vacuum Society of Japan