書誌事項
- タイトル別名
-
- The Measurement of Spectral Irradiance in Short UV Region down to 200nm
この論文をさがす
説明
A new method of spectroradiometry for short UV region is developed and spectral irradiance of deuterium lamps are determined down to 200 nm.<BR>This method is based upon the flat spectral sensitivity of thermal detector. The spectral irradiance from the source is determined by the detector without measuring spectral transmittance of the monochromator.<BR>The principle of this method is summarized as follows:<BR>(1) The quasi-monochromatic radiation from the first monochromator, with a compact xenon arc, is calibrated with the standard thermopile.<BR>(2) The calibrated quasi-monochromatic radiation and continuous spectrum of test source are received alternately the reflecting diffuser, which forms an entrance port of the second monochromater.<BR>(3) The second monochromator constitutes the spectroradiometor. The spectral irradiance of test source is determined by comparing of calibrated quasi-monochromatic radiation with test source.<BR>The results on deuterium lamp are with agreed well previous values in the overlapping rang from 225 nm to 200 nm.<BR>The aging characteristics of spectral irradiance of deuterium and hydrogen lamps are also examined.
収録刊行物
-
- 照明学会誌
-
照明学会誌 59 (9), 462-467, 1975
一般社団法人 照明学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001204077463296
-
- NII論文ID
- 130003924776
-
- ISSN
- 1349838X
- 21851506
- 00192341
- http://id.crossref.org/issn/00192341
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
- OpenAIRE
-
- 抄録ライセンスフラグ
- 使用不可