Domestic Intercomparison of Relative Spectral

Bibliographic Information

Other Title
  • 光検出器分光応答度の国内比較
  • シリコンフォトダイオードによる
  • Responsivity of Silicon Photodiodes

Abstract

An intercomparison of relative spectral responsivities of silicon photodiodes was carried out among four laboratories, ETL (Electrotechnical Laboratory), Saitama University, Hamamatsu Photonics Co., Ltd. and Matsushita Electric Industrial Co., Ltd.<BR>The photodiodes were calibrated in the 250-800nm range using each laboratory's spectral responsivity standards. As a result, the relative spectral responsivities of the photodiodes calibrated by each laboratory agreed within 1.1% standard deviation in the 400-800nm, 2.2% in 315-400nm, 3.9% in 280-315nm and 4.1% in 250-280nm. Therefore, it is confirmed that the 4 laboratories have sufficient accuracy for calibration of the relative spectral responsivity of the detector in the ultraviolet and visible wavelength region.

Journal

Citations (1)*help

See more

Details 詳細情報について

Report a problem