Evaluation of Stat Assay of Serum Digoxin Concentration by Radioimmunoassay and Its Application for Digoxin Regimen Based on Pharmacokinetics

Bibliographic Information

Other Title
  • ラジオイムノアッセイによる血清ジゴキシン濃度の測定
  • ラジオイムノアッセイ ニヨル ケッセイ ジゴキシン ノウド ノ ソクテイ ジン
  • ―迅速測定法の検討と薬物動態理論に基づく治療設計への応用―

Search this article

Abstract

Serum digoxin concentrations measured by stat RIA (phadebas digoxin RIA kit) correlated well with results obtained by complete assays. The result of stat assay can be reported within 1 hour, measuring one or more samples together with 2 standard samples in duplicates. Precise measurement can be expected with serum digoxin concentration over 0.5 ng/ml.<BR>The stat assay allows to apply the theory of pharmacokinetics for the estimation of digoxin concentration at steady states measuring minimum digoxin concentration (Cn (min) ) on the 3rd to 6th day after the start of digoxin therapy. The estimated serum digoxin levels were well agreed with measured values with the difference ranging from Y.6 to 8.6% in CV.<BR>The method is usefull for the planning and assessment of appropriate digoxin regimen.

Journal

  • RADIOISOTOPES

    RADIOISOTOPES 29 (9), 432-437, 1980

    Japan Radioisotope Association

Details 詳細情報について

Report a problem