Application of Low Background Liquid Scintillation Counting Method to Pharmacy:I. Measurement of Ultra-low Level 14C Activity in Urine
-
- BABA Shigeo
- School of Pharmacy, Tokyo University of Pharmacy and Life Sciences
-
- HORIE Masanobu
- School of Pharmacy, Tokyo University of Pharmacy and Life Sciences
-
- YANAGI Mashiho
- School of Pharmacy, Tokyo University of Pharmacy and Life Sciences
-
- KATO Yuka
- Bioscience System Engineering Section, Measuring System Engineering Dept., ALOKA Co., Ltd.
-
- YOSHIMURA Tomoyuki
- Bioscience System Engineering Section, Measuring System Engineering Dept., ALOKA Co., Ltd.
Bibliographic Information
- Other Title
-
- 低バックグラウンド液体シンチレーション計数法の薬学への応用 第1報 尿中の極低レベル`14´Cの測定
- 低バックグラウンド液体シンチレーション計数法の薬学への応用(第1報)尿中の極低レベル14Cの測定
- テイバックグラウンド エキタイ シンチレーション ケイスウホウ ノ ヤクガク エ ノ オウヨウ ダイ1ポウ ニョウ チュウ ノ ゴクテイ レベル 14C ノ ソクテイ
Search this article
Description
Utilization of a low background liquid scintillation counter(Low BG LSC)has been proposed for quantification of ultra-low level 14C in urine. The detector of this system is composed of a central counter and a guard counter, both of which are connected electronically with an anti-coincidence circuit. By this device the background(BG) counts can be decreased approximately to one third in comparison with a conventional liquid scintillation counter(LSC). Three photomultiplier tubes are used in Low BG LSC. High counting efficiency is also achieved by this improvement. 5mL of urine spiked with 14C is counted for 100 minutes and analyzed with the optimal window of multichannel analyzer. The limit of quantification(LOQ) that is defined as radioactivity giving10times standard deviation of BG sample is estimated to be 8.5 mBq per mL urine which is two times improved in comparison with accelerator mass spectrometry. The linearities of sub-becquerel 14C counted for 10 and 100 minutes by use of Low BG LSC are compared with those of LSC. The figure of merit is four times improved in Low BG LSC.
Journal
-
- RADIOISOTOPES
-
RADIOISOTOPES 58 (5), 169-178, 2009
Japan Radioisotope Association
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001204157876608
-
- NII Article ID
- 130004511523
- 10024781724
-
- NII Book ID
- AN00351589
-
- COI
- 1:CAS:528:DC%2BD1MXntVGmtb0%3D
-
- ISSN
- 18844111
- 00338303
-
- NDL BIB ID
- 10316875
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL Search
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed