超広角走査レーザ検眼鏡Optos200Tx™による眼底写真撮影法

書誌事項

タイトル別名
  • Performance of Fundus Photography Using the Ultra-Widefield Scanning Laser Ophthalmoscope Optos 200Tx™

抄録

Purpose: To compare the performance of the ultra-widefield scanning laser ophthalmoscope Optos 200Tx™ (the Optos 200Tx) and the conventional non-mydriatic auto fundus camera with a 45° picture angle (the NIDEK AFC-230).<BR>Subjects and Methods: One normal eye and one eye with diabetic retinopathy were examined by the Optos 200Tx and the NIDEK AFC-230. The methods used for photography and the size of the peripheral area that could be captured were compared between these two devices. The minimal pupil size that could be captured by the Optos 200Tx was also evaluated.<BR>Results: Because the subject's face had to be covered during the photography by the Optos 200Tx, the subject might feel less comfortable. In addition, the eyelash could be accidentally included in the image and the quality of the image might need better improvement. While the Nidek AFC-230 obtained conventional panoramic images of 100°-120° in picture angle within several tries, the Optos 200Tx could obtain images of 200° in a single capture even in pupils as small as 1.9 mm in diameter.<BR>Conclusion: As compared with the Nidek AFC-230, the Optos 200Tx could obtain ultra-widefield fundus images of eyes with small pupils in one capture. However, special care is needed to prevent the eyelashes from being included in the image. A brief explanation regarding the possible discomfort during the photography might be necessary.

収録刊行物

参考文献 (3)*注記

もっと見る

詳細情報 詳細情報について

  • CRID
    1390001204245661184
  • NII論文ID
    130004552639
  • DOI
    10.4263/jorthoptic.041f122
  • ISSN
    18839215
    03875172
  • 本文言語コード
    ja
  • データソース種別
    • JaLC
    • Crossref
    • CiNii Articles
  • 抄録ライセンスフラグ
    使用不可

問題の指摘

ページトップへ