Trace Analysis of Released Metallic Ions in Static Immersion Test for Characterlization of Metallic Biomaterials
-
- Ashino Tetsuya
- Institute for Materials Research, Tohoku University
-
- Ohtsu Naofumi
- Institute for Materials Research, Tohoku University
-
- Wagatsuma Kazuaki
- Institute for Materials Research, Tohoku University
書誌事項
- タイトル別名
-
- Trace Analysis of Released Metallic Ions in Static Immersion Test for Characterization of Metallic Biomaterials
この論文をさがす
抄録
For characterization of corrosion resistance of metallic biomaterials, determination of trace amounts of metallic ions released from the materials in using static immersion into simulated body fluids (SBF) were investigated. In using a H2SO4-fume pre-treatment method, sensitive, precise and accurate determination of the trace metallic elements in SBFs could be performed by ICP-OES. For accurate analysis, it was necessary to employ a matrix-matched solution for the calibration. Moreover, usage of a PFA vessel in the static immersion test was recommended for the prevention of contamination. Thus, it was possible to determine μg dm−3 (ppb) order of elements in SBFs and evaluate nano-gram order of the released metallic ions.
収録刊行物
-
- MATERIALS TRANSACTIONS
-
MATERIALS TRANSACTIONS 49 (6), 1342-1345, 2008
公益社団法人 日本金属学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390001204249420928
-
- NII論文ID
- 10021147557
-
- NII書誌ID
- AA1151294X
-
- ISSN
- 13475320
- 13459678
-
- HANDLE
- 10097/52264
-
- NDL書誌ID
- 9519269
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- IRDB
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可