Conically Shaped Chromatographic Column.

この論文をさがす

抄録

We investigated a column with varying physical properties. A change in velocity was induced by a continuously varying the diameter inside the column. Analytical dimensions of the column were used.The diameter was continuously reduced from the analytical one (4.0 mm) to a diameter typical for narrow-bore columns (2.5 mm). Our interest was focused on the efficiency of a conical column and on the mobile phase velocity in it. The results obtained with the conical column were compared to those obtained with the conventional analytical column and the narrow column. The diameter of the narrow column (3.3 mm) was the same as the average diameter of the conical column. The advantage of the conical column over the narrow column was the greater loadability of the conical column and better dynamic efficiency. Compared to the conventional analytical column, a higher linear velocity at the same flow rate was achieved. Therefore, the retention times were shorter and consumption of the mobile phase was smaller.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 15 (3), 233-240, 1999

    社団法人 日本分析化学会