Design and Application of Piezoelectric Quartz Crystal-based Immunoassay.
-
- SU Xiaodi
- Institute of Material Research and Engineering
-
- CHEW Fook Tim
- Paediatrics Department, Medical Faculty, National University of Singapore
-
- LI Sam F. Y.
- Institute of Material Research and Engineering
この論文をさがす
抄録
This paper describes some basic test schemes of piezoelectric quartz crystal microbalance (QCM)-based immunoassay. These include one-step direct assay, one-step indirect assay, two-step sandwiched assay, displacement assay, and mass-amplified assay. Applications of these designs in the fields of clinic diagnosis, food analysis, environmental analysis, and veterinary diagnosis are reviewed. In addition, some critical issues of QCM immunoassay are also dealt with, including the evaluation of liquid phase and gas phase detection, and the biomolecules’ coating techniques, non-specific adsorption, and regeneration of used crystals.
収録刊行物
-
- Analytical Sciences
-
Analytical Sciences 16 (2), 107-114, 2000
社団法人 日本分析化学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390001204255554176
-
- NII論文ID
- 130004440208
- 10004704899
-
- NII書誌ID
- AA10500785
-
- COI
- 1:CAS:528:DC%2BD3cXht1yltLw%3D
-
- ISSN
- 13482246
- 09106340
-
- NDL書誌ID
- 5256161
-
- Web Site
- http://id.ndl.go.jp/bib/5256161
- https://ndlsearch.ndl.go.jp/books/R000000004-I5256161
- https://link.springer.com/content/pdf/10.2116/analsci.16.107.pdf
- https://link.springer.com/article/10.2116/analsci.16.107/fulltext.html
- http://www.jstage.jst.go.jp/article/analsci/16/2/16_2_107/_pdf
- https://search.jamas.or.jp/link/ui/2000166202
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可