Isotope Dilution Analysis of Selenite and Selenate in Natural Water Using Microwave-Induced Nitrogen Plasma Mass Spectrometry

DOI 機関リポジトリ PDF Web Site Web Site ほか2件をすべて表示 一部だけ表示 被引用文献1件 参考文献56件 オープンアクセス

この論文をさがす

抄録

Isotope dilution analysis of the sub-μg l-1 levels of selenite and selenate in natural water samples by microwave-induced nitrogen plasma mass spectrometry (MIP-MS) was performed. An appropriate amount of a spike solution containing 78Se-selenite and 78Se-selenate was added to the natural water sample to be analyzed. Both analytes in the water were then concentrated simultaneously by passing the sample through a column that was filled with an anionic exchange resin. After the concentration process, all of the selenite and some of the selenate on the resin were eluted by 0.03 M nitric acid. The residual selenate was eluted by 0.13 M nitric acid. The eluted sample solutions were injected into MIP-MS, and isotope dilution analyses were carried out. Selenite and selenate concentrations as low as 0.01 μg l-1 in the natural water sample were successfully determined by the proposed method.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 19 (10), 1359-1363, 2003

    社団法人 日本分析化学会

被引用文献 (1)*注記

もっと見る

参考文献 (56)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘

ページトップへ