Sulfate-selective Electrode Based on a Complex of Copper

この論文をさがす

抄録

A new sulfate-selective electrode based on the complex N,N′-bis(2-amino-1-oxo-phenelenyl)phenylenediamine copper(II) (CuL) as the membrane carrier was developed. The electrode exhibited a good Nernstian slope of -29.5 ± 0.5 mV/decade and a linear range of 1.0 × 10-7 - 1.0 × 10-1 M for sulfate. The limit of detection was 1.0 × 10-8 M. It has a fast response time of 10 s and can be used for more than three months. The selective coefficients were determined by the fixed interference method (FIM). The electrode could be used in the pH range 3.5 - 8.0. It was employed as an indicator electrode for direct determination of sulfate in pharmacy and cement samples.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 22 (5), 673-678, 2006

    社団法人 日本分析化学会

参考文献 (64)*注記