Measurement of Circular Dichroism Spectra of Liquid/Liquid Interface by Centrifugal Liquid Membrane Method

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 被引用文献9件 参考文献27件

この論文をさがす

説明

The direct measurement of the circular dichroism (CD) spectra of liquid/liquid interface has been achieved for the first time by the centrifugal liquid membrane (CLM) method combined with a conventional CD spectropolarimetry. In the sample chamber of the CD spectropolarimeter, a cylindrical glass cell containing small amounts of organic and aqueous phases was rotated at ca. 7000 rpm to generate a two-phase liquid membrane with a high specific interfacial area. The CD spectra of the J-aggregate of protonated 5,10,15,20-tetraphenylporphyrin formed at the toluene/sulfuric acid interface in the rotating cell have been measured. The results demonstrated the novelty and advantages of this method.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 20 (11), 1489-1491, 2004

    社団法人 日本分析化学会

被引用文献 (9)*注記

もっと見る

参考文献 (27)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘